Author:
Lehner Sebastian,Graner Frank Philipp,Fischer Maximilian,Ilhan Harun,Bartenstein Peter,Todica Andrei
Abstract
Abstract
Background
Due to partly conflicting studies, further research is warranted with the QGS software package, with regard to the performance of gated FDG PET phase analysis as compared to gated MPS as well as the establishment of possible cut-off values for FDG PET to define dyssynchrony.
Methods
Gated MPS and gated FDG PET datasets of 93 patients were analyzed with the QGS software. BW, Phase SD, and Entropy were calculated and compared between the methods. The performance of gated PET to identify dyssynchrony was measured against SPECT as reference standard. ROC analysis was performed to identify the best discriminator of dyssynchrony and to define cut-off values.
Results
BW and Phase SD differed significantly between the SPECT and PET. There was no significant difference in Entropy with a high linear correlation between methods.
There was only moderate agreement between SPECT and PET to identify dyssynchrony. Entropy was the best single PET parameter to predict dyssynchrony with a cut-off point at 62%.
Conclusion
Gated MPS and gated FDG PET can assess LVMD. The methods cannot be used interchangeably. Establishing reference ranges and cut-off values is difficult due to the lack of an external gold standard. Further prospective research is necessary.
Funder
Ludwig-Maximilians-Universität München
Publisher
Springer Science and Business Media LLC
Subject
Cardiology and Cardiovascular Medicine,Radiology, Nuclear Medicine and imaging
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献