Abstract
Abstract
Purpose
Uncemented stem migration analysis by EBRA-FCA (Einzel-Bild-Roentgen Analyse, Femoral Component Analyse) has been seen to be a good predictive indicator for early implant failure. In this study, we investigated the migration behavior of a cementless metaphyseal-anchored press-fit stem after 4-year follow-up.
Methods
Applying a retrospective study design, we reviewed all consecutive patients who between 2012 and 2017 received a cementless Accolade II press-fit stem at our Department. We reviewed medical histories and performed radiological measurements using EBRA-FCA software. EBRA-FCA measurements and statistical investigations were performed by two independent investigators.
Results
A total of 102 stems in 91 patients (female 60; male 31) fulfilled our inclusion criteria. Mean age at surgery was 66.2 (range 24.3–92.6) years. EBRA migration analysis showed a mean subsidence of 1.4 mm (range 0.0–12.0) at final follow-up. The angle between stem and femur axis was 0.5° (range 0.0°–2.8°) after 48 months. No correlations between gender or Dorr types and subsidence were found (p > 0.05). A body mass index > 30 kg/m2 showed a significant increase in stem subsidence within the first 6 (p = 0.0258) and 12 months (p = 0.0466) postoperative.
Conclusions
Migration pattern of the metaphyseal-anchored stem and a low subsidence rate at final follow-up may predict a good long-term clinical result.
Trial registration
Number: 20181024-1875.
Funder
University of Innsbruck and Medical University of Innsbruck
Publisher
Springer Science and Business Media LLC
Subject
Orthopedics and Sports Medicine,General Medicine,Surgery,Surgery
Cited by
3 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献