Long-term survivorship of an exchangeable-neck hip prosthesis with a Ti-alloy/Ti-alloy neck–stem junction

Author:

Baleani MassimilianoORCID,Toni AldoORCID,Ancarani CristinaORCID,Stea SusannaORCID,Bordini BarbaraORCID

Abstract

Abstract Introduction Breakage of exchangeable-neck (EN) and adverse local tissue reactions (ALTRs) to neck–stem junction (NSJ) damage products are responsible for increasing the revision rate of EN hip prostheses. We investigated the survivorship of an EN hip prosthesis including a NSJ with both components made of titanium alloy (Ti-alloy/Ti-alloy) to assess whether, and to what extent, EN breakage and NSJ damage affected implant survivorship. Materials and methods Using data from a hip replacement registry, we determined survivorship of 2857 EN prostheses. Long-offset configurations of head and EN were implanted in heavy (> 90 kg) patients only in 23 hips. We investigated under which conditions EN breakages or ALTRs occurred. We also measured titanium (Ti) and vanadium (V) blood concentrations in 24 patients with a unilateral well-working prosthesis. Results The 17-year survival rates for any reason and aseptic loosening of any component were 88.9% (95%CI 87.5–90.1; 857 hips at risk) and 96.9% (95%CI 96.0–97.6), respectively. There were two cases of EN breakage and one case of ALTR (metallosis), due to rim-neck impingement, out of 276 revisions. After an average period of 9.8 years (range 7.8–12.8 years), the maximum Ti and V blood concentrations in patients with a well-working prosthesis were 5.0 µg/l and 0.16 µg/l, respectively. Conclusion The present incidence of EN breakage or ALTR is lower than those reported in other studies evaluating EN hip prosthesis survivorship. This study suggests that (i) the risk of EN breakage is reduced by limiting the use of long-offset configurations in heavy patients and (ii) Ti-alloy/Ti-alloy NSJ damage products do not promote ALTR nor significantly alter the rate of implant loosening. Since design decisions and implant configuration determine the NSJ strength, the NSJ strength in working conditions must be thoroughly investigated to proper define the clinical indications for any EN design.

Funder

Italian Ministry of Health

Publisher

Springer Science and Business Media LLC

Subject

Orthopedics and Sports Medicine,General Medicine,Surgery,Surgery

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3