STRYDE versus PRECICE magnetic internal lengthening nail for femur lengthening

Author:

Galal SherifORCID,Shin Jonggu,Principe Peter,Khabyeh-Hasbani Nathan,Mehta Rena,Hamilton Amber,Rozbruch S. Robert,Fragomen Austin T.

Abstract

Abstract Introduction Magnetic internal lengthening nails (MILNs) have been used for femoral lengthening to avoid complications associated with external fixation. The titanium version of the MILN (PRECICE®) has been in use since 2011 but had limitations (50–75 lb) in post-operative weight bearing. A new stainless-steel version of the MILN (STRYDE®) allows 150–250 lb of post-operative weight bearing. The aim is to compare the outcomes of using these two different MILNs for both unilateral and bilateral femoral lengthening. Methods A single-center, retrospective cohort study was conducted in which patients’ records were reviewed from the period from January 2017 to March 2020. A total of 66 femoral lengthening procedures were included in the study and were divided into two groups: STRYDE® group (30 femora) and PRECICE® group (36 femora). Outcomes assessed were the 6-months post-operative Limb Deformity-Scoliosis Research Society (LD-SRS) Score, adjacent joint range of motion (ROM), average distraction rate, bone healing index (BHI), and complications. Results No statistically significant difference was found between the two groups in regard to the (LD-SRS) score, hip ROM, or knee ROM. Statistically significant differences were found between the two groups in regard to BHI (average of 0.84 months/cm and 0.67 months/cm for STRYDE® and PRECICE®, respectively) and distraction rate (average of 0.6 mm/day and 0.9 mm/day for STRYDE® and PRECICE®, respectively). No mechanical nail complications were reported in the STRYDE® group compared to three events of nail failure in the PRECICE® group. One femur in the PRECICE® group needed BMAC injection for delayed healing compared to four femurs in the STRYDE® group. Conclusion The STRYDE® MILN yields comparable functional results to those of PRECICE® MILN and shows fewer mechanical nail complications. However, STRYDE® MILN requires a slower distraction rate and yields slower healing (larger BHI). Level of evidence Level III, Therapeutic study.

Publisher

Springer Science and Business Media LLC

Subject

Orthopedics and Sports Medicine,General Medicine,Surgery,Surgery

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3