Abstract
Abstract
Introduction
As endpoint of a prospective multicenter 10-year documentation using the Columbus system, this evaluation carried out results of clinical scores (Knee Society Score and Oxford Knee Score), an evaluation of radiological imaging, survival rates and a collection of complication statistics.
Materials and methods
There was a multicenter prospective recruitment of consecutive patients with the indication for total knee replacement (TKR). Preoperatively and 10 years after implantation, clinical scores, range of motion and radiological imaging was performed. During this period, a detailed documentation of complications was made.
Results
A total of 210 patients were recruited in 5 centers. 140 patients were available for endpoint examination 10 years after surgery. A survival rate of 96.6% (CI 95%) for the implanted Columbus prostheses after 10 years was demonstrated. Cumulative KSS showed an improvement of 75.3 (± 38.1) points and was highly significant (p < 0.0001, t test). The average functional improvement in the Oxford score was 20.6 (± 9.5) points and was also highly significant (p < 0.0001). The ROM improved from 106.5° (± 20.5) preoperatively to 116.0° (± 11.5) (p < 0.0001, t test). There were no implant-related complications as well as no new complication documented between 5- and 10-year follow-up.
Conclusions
The endpoint analysis after an observation period of 10 years provided good clinical and radiographic results. In particular, an excellent survival rate of 96.6% after 10 years was demonstrated. The data published in this study are the first to be available in a prospective multicenter study on this system, which leads to a high level of clinical significance.
Funder
Aesculap Implant Systems
Albert-Ludwigs-Universität Freiburg im Breisgau
Publisher
Springer Science and Business Media LLC
Subject
Orthopedics and Sports Medicine,General Medicine,Surgery,Surgery
Reference32 articles.
1. The Australian Orthopaedic Association National Joint Replacement Registry—Annual Report 2016. 2016.
2. The Australian Orthopaedic Association National Joint Replacement Registry, Annual Report 20192019.
3. Beyer F, Lutzner C, Kirschner S, Lutzner J (2016) Midterm results after coated and uncoated TKA: A Randomized Controlled Study. Orthopedics 39:S13-17
4. Bhatt H, Rambani R, White W, Chakrabarty G (2012) Primary total knee arthroplasty using the P.F.C Sigma(R)-rotating platform cruciate retaining endoprosthesis—a 6 year follow up. Knee 19:856–859
5. Cawley DT, Kelly N, McGarry JP, Shannon FJ (2013) Cementing techniques for the tibial component in primary total knee replacement. Bone Joint J. 95:295–300
Cited by
2 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献