In vivo study to assess fat embolism resulting from the Reamer-Irrigator-Aspirator 2 system compared to a novel aspirator-based concept for intramedullary bone graft harvesting

Author:

Laubach MarkusORCID,Bessot Agathe,Saifzadeh Siamak,Savi Flavia Medeiros,Hildebrand Frank,Bock Nathalie,Hutmacher Dietmar W.,McGovern Jacqui

Abstract

Abstract Introduction Fat embolism (FE) following intramedullary (IM) reaming can cause severe pulmonary complications and sudden death. Recently, a new harvesting concept was introduced in which a novel aspirator is used first for bone marrow (BM) aspiration and then for subsequent aspiration of morselized endosteal bone during sequential reaming (A + R + A). In contrast to the established Reamer-Irrigator-Aspirator (RIA) 2 system, the new A + R + A concept allows for the evacuation of fatty BM prior to reaming. In this study, we hypothesized that the risk of FE, associated coagulopathic reactions and pulmonary FE would be comparable between the RIA 2 system and the A + R + A concept. Materials and methods Intramedullary bone graft was harvested from intact femora of 16 Merino sheep (age: 1–2 years) with either the RIA 2 system (n = 8) or the A + R + A concept (n = 8). Fat intravasation was monitored with the Gurd test, coagulopathic response with D-dimer blood level concentration and pulmonary FE with histological evaluation of the lungs. Results The total number and average size of intravasated fat particles was similar between groups (p = 0.13 and p = 0.98, respectively). D-dimer concentration did not significantly increase within 4 h after completion of surgery (RIA 2: p = 0.82; A + R + A: p = 0.23), with an interaction effect similar between groups (p = 0.65). The average lung area covered with fat globules was similar between groups (p = 0.17). Conclusions The use of the RIA 2 system and the novel A + R + A harvesting concept which consists of BM evacuation followed by sequential IM reaming and aspiration of endosteal bone, resulted in only minor fat intravasation, coagulopathic reactions and pulmonary FE, with no significant differences between the groups. Our results, therefore, suggest that both the RIA 2 system and the new A + R + A concept are comparable technologies in terms of FE-related complications.

Funder

ARC Training Centre for Multiscale 3D Imaging, Modelling and Manufacturing

Queensland University of Technology

Publisher

Springer Science and Business Media LLC

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3