Abstract
Abstract
Purpose
To assess the implementation and outcome quality of the Freiburg Acuity VEP methodology (Bach et al. in Br J Ophthalmol 92:396–403, 2008) on the Diagnosys Espion Profile and E3 electrophysiology systems.
Methods
We recorded visual evoked potentials (VEPs) from both eyes of 24 participants, where visual acuity (VA) was either full or reduced with scatter foils to approximately 0.5 and 0.8 LogMAR, resulting in a total of 144 recordings. Behavioral VA was measured in each case under the same conditions using the Freiburg Acuity Test (FrACT); VEP-based acuity was assessed with the “heuristic algorithm,” which automatically selects points for regression to zero amplitude.
Results
Behavioral VA ranged from − 0.2 to 1.0 LogMAR. The fully automatic heuristic VEP algorithm resulted in 8 of 144 recordings (6%) that were scored as “no result.” The other 136 recordings (94%) had an outcome of − 0.20 to 1.3 LogMAR (which corresponds to a range of 20/12.5–20/400, or 6/3.8–6/120, in Snellen ratios; or 1.6–0.1 in decimal acuity). The heuristic VEP algorithm agreed with the behavioral VA to within ± 0.31 LogMAR (95% limits of agreement), which is equivalent to approximately three lines on a VA chart.
Conclusions
The successful implementation of the Freiburg Acuity VEP “heuristic algorithm” on a commercial system makes this capability available to a wider group of users. The limits of agreement of ± 0.31 LogMAR are close to the original implementation at the University of Freiburg and we believe are clinically acceptable for a fully automatic, largely objective assessment of visual acuity.
Publisher
Springer Science and Business Media LLC
Subject
Physiology (medical),Sensory Systems,Ophthalmology
Reference19 articles.
1. Bach M, Maurer JP, Wolf ME (2008) Visual evoked potential-based acuity assessment in normal vision, artificially degraded vision, and in patients. Br J Ophthalmol 92:396–403.
https://doi.org/10.1136/bjo.2007.130245
2. Almoqbel F, Leat SJ, Irving E (2008) The technique, validity and clinical use of the sweep VEP. Ophthalmic Physiol Opt 28:393–403.
https://doi.org/10.1111/j.1475-1313.2008.00591.x
3. Meigen T, Bach M (1999) On the statistical significance of electrophysiological steady-state responses. Doc Ophthalmol 98:207–232
4. Bach M, Joost W (1989) VEP vs spatial frequency at high contrast: subjects have either a bimodal or single-peaked response function. In: Kulikowski J, Dickinson C, Murray I (eds) Seeing contour and colour. Pergamon Press, Oxford, pp 478–484
5. Wenner Y, Heinrich SP, Beisse C et al (2014) Visual evoked potential-based acuity assessment: overestimation in amblyopia. Doc Ophthalmol 128:191–200.
https://doi.org/10.1007/s10633-014-9432-3
Cited by
5 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献