Abstract
Abstract
Purpose
The aim of this study was to investigate the feasibility and reliability of transcutaneous ultrasound for the detection of complications after cochlear implantation.
Methods
In a single center retrospective cohort study, 115 consecutive cases of suspected complications after cochlear implantation (intervention group) were examined. The rate of pathologic ultrasound findings for specific leading symptoms and diagnoses was compared to a control group comprising twenty consecutive cochlear implants in symptom-free patients.
Results
Diagnostic ultrasound showed distinctly more pathologic findings in the intervention group (n = 67; 58.3%; p < 0.001) compared to the control group (n = 1; 5%). Ultrasound revealed significantly more pathologic findings in haematoma or seroma around the implant (n = 17; 100%; p < 0.001; ϕ = 0.94) and magnet dislocation (n = 44; 97.7%; p < 0.001; ϕ = 0.92) confirmed by a strong effect. Ultrasound examination showed a medium to high effect size in patients presenting with local infections (n = 3; 21.4%; p = 0.283; ϕ = 0.25) and skin flap oedema (n = 2; 50%; p = 0.061; ϕ = 0.51). In contrast, ultrasound examinations displayed a low effect size in undefined cephalgia (0%; p = 0.444; ϕ = 0.17) and device malfunction or failure (0%; p > 0.999; ϕ = 0.13).
Conclusion
Transcutaneous ultrasound can be advocated as a feasible and effective method in the diagnostic work-up of magnet dislocation and haematoma or seroma around the implant following cochlear implantation. Contrary, ultrasound findings can be expected to be inconspicuous in patients presenting with undefined cephalgia and device malfunction or failure.
Funder
Friedrich-Alexander-Universität Erlangen-Nürnberg
Publisher
Springer Science and Business Media LLC
Subject
General Medicine,Otorhinolaryngology
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献