Abstract
Abstract
Introduction
Bone conduction devices (BCD) are effective for hearing rehabilitation in patients with conductive and mixed hearing loss or single-sided deafness. Transcutaneous bone conduction devices (tBCD) seem to lead to fewer soft tissue complications than percutaneous BCDs (pBCD) but have other drawbacks such as MRI incompatibility and higher costs. Previous cost analyses have shown a cost advantage of tBCDs. The purpose of this study is to compare long-term post-implantations costs between percutaneous and transcutaneous BCDs.
Materials and methods
Retrospective data from 77 patients implanted in a tertiary referral centre with a pBCD (n = 34), tBCD (n = 43; passive (tpasBCD; n = 34) and active (tactBCD; n = 9) and a reference group who underwent cochlear implantation (CI; n = 34), were included in a clinical cost analysis. Post-implantation costs were determined as the sum of consultation (medical and audiological) and additional (all post-operative care) costs. Median (cumulative) costs per device incurred for the different cohorts were compared at 1, 3 and 5 years after implantation.
Results
After 5 years, the total post-implantation costs of the pBCD vs tpasBCD were not significantly different (€1550.7 [IQR 1174.6–2797.4] vs €2266.9 [IQR 1314.1–3535.3], p = 0.185), nor was there a significant difference between pBCD vs tactBCD (€1550.7 [1174.6–2797.4] vs €1428.8 [1277.3–1760.4], p = 0.550). Additional post-implantation costs were significantly highest in the tpasBCD cohort at all moments of follow-up.
Conclusion
Total costs related to post-operative rehabilitation and treatments are comparable between percutaneous and transcutaneous BCDs up to 5 years after implantation. Complications related to passive transcutaneous bone conduction devices appeared significantly more expensive after implantation due to more frequent explantations.
Publisher
Springer Science and Business Media LLC
Subject
General Medicine,Otorhinolaryngology
Reference25 articles.
1. Tjellstrom A, Lindstrom J, Hallen O, Albrektsson T, Branemark PI (1981) Osseointegrated titanium implants in the temporal bone A clinical study on bone-anchored hearing aids. Am J Otol 2(4):304–310
2. Hakansson B, Liden G, Tjellstrom A, Ringdahl A, Jacobsson M, Carlsson P et al (1990) Ten years of experience with the Swedish bone-anchored hearing system. Ann Otol Rhinol Laryngol Suppl 151:1–16
3. Snik AF, Mylanus EA, Proops DW, Wolfaardt JF, Hodgetts WE, Somers T et al (2005) Consensus statements on the BAHA system: where do we stand at present? Ann Otol Rhinol Laryngol Suppl 195:2–12
4. Lagerkvist H, Carvalho K, Holmberg M, Petersson U, Cremers C, Hultcrantz M (2020) Ten years of experience with the Ponto bone anchored hearing system—a systematic literature review. Clin Otolaryngol 45:667–680
5. Dun CA, Faber HT, de Wolf MJ, Mylanus EA, Cremers CW, Hol MK (2012) Assessment of more than 1,000 implanted percutaneous bone conduction devices: skin reactions and implant survival. Otol Neurotol 33(2):192–198