Abstract
Abstract
Purpose
Hearing performance data was collected from a large heterogeneous group of subjects implanted with the Cochlear™ Nucleus® CI532 with Slim Modiolar Electrode, for the purposes of postmarket clinical follow-up. Data was analysed for factors which may predict postoperative speech recognition scores.
Methods
Data was collected retrospectively from five German clinics for 159 subjects from March 2017 to August 2018. Hearing thresholds and recognition scores for monosyllabic words in quiet and sentences in noise were measured preoperatively and at 3 and 6 months postoperatively.
Results
There was a mean gain of 44% points (95% CI 39–49%) at 6 months in monosyllable scores in quiet for implanted ears. Preoperative hearing thresholds in implant ears increased systematically with decreasing age; however, younger subjects had better baseline monosyllable scores with hearing aids compared with older subjects. Baseline performance alone explained 14% of the variation in postoperative scores. Residual hearing was preserved on average to within 22 dB at 250 Hz and 30 dB at 500 Hz of preoperative levels.
Conclusions
In a large and varied cohort of routinely treated hearing-impaired adults, speech recognition with the CI532 for German monosyllabic words in quiet at 6 months was equivalent to performance reported at one year or more in other published studies. Although younger subjects had poorer preoperative pure-tone thresholds, they had better preoperative word recognition scores compared with older subjects, and also had higher post implant scores. Further research is required to identify if this phenomenon is just applicable to German health system assessment and referral practices.
Publisher
Springer Science and Business Media LLC
Subject
General Medicine,Otorhinolaryngology
Reference45 articles.
1. Aschendorff A, Briggs R, Brademann G, Helbig S, Hornung J, Lenarz T, Marx M, Ramos A, Stöver T, Escudé B, James CJ (2017) Clinical investigation of the nucleus slim modiolar electrode. Audiol Neurootol 22(3):169–179
2. Shaul C, Dragovic AS, Stringer AK, O'Leary SJ, Briggs RJ (2018) Scalar localisation of peri-modiolar electrodes and speech perception outcomes. J Laryngol Otol 29:1–7
3. McJunkin JL, Durakovic N, Herzog J, Buchman CA (2018) Early outcomes with a slim, modiolar cochlear implant electrode array. Otol Neurotol 39(1):e28–e33
4. Carlson ML, Driscoll CL, Gifford RH, Service GJ, Tombers NM, Hughes-Borst BJ, Neff BA, Beatty CW (2011) Implications of minimizing trauma during conventional cochlear implantation. Otol Neurotol. 32(6):962–968
5. Dalbert A, Huber A, Baumann N, Veraguth D, Roosli C, Pfiffner F (2016) Hearing preservation after cochlear implantation may improve long-term word perception in the electric-only condition. Otol Neurotol 37(9):1314–1319