Abstract
Abstract
Purpose
The Vibrant Soundbridge (VSB) was introduced in 1996, and the fourth generation of the audio processor recently released. This clinical study evaluates the audiological performance and subjective satisfaction of the new SAMBA 2 audio processor compared to its predecessor, SAMBA.
Method
Fifteen VSB users tested both audio processors for approximately 3 weeks. Air conduction and bone conduction thresholds and unaided and aided sound field thresholds were measured with both devices. Speech performance in quiet (Freiburg monosyllables) and noise (OLSA) was evaluated as well as subjective listening effort (ACALES) and questionnaire outcomes (SSQ12 and APSQ). In addition, data from 16 subjects with normal hearing were gathered on sound field tests and ACALES.
Results
Both audio processors showed substantial improvement compared to the unaided condition. The SAMBA and SAMBA 2 had comparable performance in sound filed thresholds, while the SAMBA 2 was significantly better in speech in quiet, speech in noise, reduced listening effort, and improved subjective satisfaction compared with the SAMBA.
Conclusion
The SAMBA 2 audio processor, compared to its predecessor SAMBA, offers improved performance throughout the parameters investigated in this study. Patients with a VSB implant would benefit from an upgrade to SAMBA 2.
Funder
Martin-Luther-Universität Halle-Wittenberg
Publisher
Springer Science and Business Media LLC
Subject
General Medicine,Otorhinolaryngology
Cited by
7 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献