Abstract
Abstract
Purpose
Early detection of tumors and their spread, particularly in lymph node illnesses, is critical for a full recovery. However, it is currently difficult due to a lack of imaging or detection devices that provide the necessary spatial depth and location information. Consequently, it would be beneficial to have a simple and cost-effective sensor device to determine the 3D position of, e.g., a lymph node in the patient’s coordinate system.
Methods
In this work, we present a concept and design for a novel semiconductor-based 3D detection system that uses inexpensive off-the-shelf components to measure gamma activity. A simple Arduino-type microcontroller calculates the 3D position of the probe based on the number of the measured pulse, the spatial sensitivity characteristics, and the known geometry of the device.
Results
The system was set up from four photodiodes (Osram BPW34), a transistor-based pre-amplifier, and a two-stage operational amplifier as the main stage. Doing so, a signal sufficient to be read by the microcontroller could be produced. The performed calculations proved that for a system consisting of at least four photodiodes, it is possible to determine precise location of a gamma radiation source.
Conclusions
After successful first experiments with a single diode, the optimal spatial arrangement of the diodes as well as their orientation will be determined to achieve a compact, cost effective yet fast, and accurate sensor device for every-day clinical application.
Publisher
Springer Science and Business Media LLC
Subject
Health Informatics,Radiology, Nuclear Medicine and imaging,General Medicine,Surgery,Computer Graphics and Computer-Aided Design,Computer Science Applications,Computer Vision and Pattern Recognition,Biomedical Engineering