Abstract
Abstract
Purpose
Automatic identification of interventional devices in X-ray (XR) fluoroscopy offers the potential of improved navigation during transcatheter endovascular procedures. This paper presents a prototype implementation of fully automatic 3D reconstruction of a cryo-balloon catheter during pulmonary vein isolation (PVI) procedures by deep learning approaches.
Methods
We employ convolutional neural networks (CNN) to automatically identify the cryo-balloon XR marker and catheter shaft in 2D fluoroscopy during PVI. Training data are generated exploiting established semiautomatic techniques, including template-matching and analytical graph building. A first network of U-net architecture uses a single grayscale XR image as input and yields the mask of the XR marker. A second network of the similar architecture is trained using the mask of the XR marker as additional input to the grayscale XR image for the segmentation of the cryo-balloon catheter shaft mask. The structures automatically identified in two 2D images with different angulations are then used to reconstruct the cryo-balloon in 3D.
Results
Automatic identification of the XR marker was successful in 78% of test cases and in 100% for the catheter shaft. Training of the model for prediction of the XR marker mask was successful with 3426 training samples. Incorporation of the XR marker mask as additional input for the model predicting the catheter shaft allowed to achieve good training result with only 805 training samples. The average prediction time per frame was 14.47 ms for the XR marker and 78.22 ms for the catheter shaft. Localization accuracy for the XR marker yielded on average 1.52 pixels or 0.56 mm.
Conclusions
In this paper, we report a novel method for automatic detection and 3D reconstruction of the cryo-balloon catheter shaft and marker from 2D fluoroscopic images. Initial evaluation yields promising results thus indicating the high potential of CNNs as alternatives to the current state-of-the-art solutions.
Funder
Bundesministerium für Bildung und Forschung
Universitätsklinikum Ulm
Publisher
Springer Science and Business Media LLC
Subject
Health Informatics,Radiology Nuclear Medicine and imaging,General Medicine,Surgery,Computer Graphics and Computer-Aided Design,Computer Science Applications,Computer Vision and Pattern Recognition,Biomedical Engineering
Reference15 articles.
1. Kurzendorfer T, Mewes PW, Maier A, Strobel N, Brost A (2016) Cryo-balloon catheter localization based on a support-vector-machine approach. IEEE Trans Med Imaging 35(8):1892–1902
2. Bourier F, Fahrig R, Wang P, Santangeli P, Kurzidim K, Strobel N, Moore T, Hinkel C, Al-Ahmad A (2014) Accuracy assessment of catheter guidance technology in electrophysiology procedures. J Cardiovasc Electrophysiol 25(1):74–83
3. Kowalewski CAB, Rodrigo M, Brodt C, Haddad F, Wang PJ, Narayan SM (2020) Novel three-dimensional imaging approach for cryoballoon navigation and confirmation of pulmonary vein occlusion. Pacing Clin Electrophysiol PACE 43(3):269–277
4. Hoffmann M, Brost A, Jakob C, Bourier F, Koch M, Kurzidim K, Hornegger J, Strobel N (2012) Semi-automatic catheter reconstruction from two views. MICCAI 2012. Lecture Notes in Computer Science. pp 584–591
5. Bourier F, Brost A, Kleinoeder A, Kurzendorfer T, Koch M, Kiraly A, Schneider HJ, Hornegger J, Strobel N, Kurzidim K (2012) Navigation for fluoroscopy-guided cryo-balloon ablation procedures of atrial fibrillation. Proc SPIE Med Imaging 2012:8316
Cited by
6 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献