Abstract
Purpose
Tether breakage is the most common complication of Vertebral Body Tethering (VBT) occurring in up to 52% of Adolescent Idiopathic Scoliosis (AIS) patients and risks continued progression and revision. Radiographical diagnosis of tether breakage is commonly defined by a 5° increase in inter-screw angle and associates breakage with loss of correction. However, the sensitivity of this method was 56% only, suggesting that tethers can break without an increase in angulation, which was supported by other studies. To our knowledge, current literature lacks a method merely focusing on the diagnosis of tether breakage radiographically that does not associate the breakages with loss of correction.
Methods
This was a retrospective review of prospectively collected data of AIS patients who underwent VBT. The “inter-screw index” is defined as the percentage increase in inter-screw distance since post-op, with ≥ 13% increase defined as tether breakage as suggested by our mechanical tests. CTs were reviewed to identify the breakages and compared with inter-screw angle and inter-screw index.
Results
94 segments from 13 CTs were reviewed, and 15 tether breakages were identified. Use of inter-screw index correctly identified 14 breakages (93%), whereas ≥ 5° increase in inter-screw angle only identified 12 breakages (80%).
Conclusion
Use of inter-screw index is proven to be more sensitive than inter-screw angle in identifying tether breakages. Therefore, we propose the use of inter-screw index to diagnose tether breakages radiographically. Tether breakages were not necessarily accompanied by a loss of segmental correction leading to an increase in inter-screw angle, especially after skeletal maturity.
Level of evidence
Level 3.
Funder
Sanming Project of Medicine in Shenzhen
Publisher
Springer Science and Business Media LLC
Subject
Orthopedics and Sports Medicine
Cited by
2 articles.
订阅此论文施引文献