Author:
Gazit Inbal,Gershevich Anna,Einan-Lifshitz Adi,Pras Eran,Barrett Graham D.,Or Lior
Abstract
Abstract
Background
The purpose of this study was to investigate the visual and refractive outcomes in patients with pseudoexfoliation (PXF) undergoing routine cataract surgery and to compare the accuracy of intraocular lens (IOL) power calculation formulae.
Methods
Retrospective case-series study from Shamir medical center, a public hospital, Israel. Medical records of patients who underwent routine cataract surgery between January 2019 and August 2021 were investigated. Postoperative visual acuity and manifest refraction were examined. The error in predicted refraction and IOL power calculation accuracy within a range of ± 0.50 to ± 1.00 diopters were compared between different IOL calculating formulae.
Results
151 eyes of 151 patients ages 73.9 ± 7.1 years were included in this study- 58 eyes in the PXF group and 93 eyes in the control group. The mean absolute error (MAE) for the BUII formula was 0.63D ± 0.87 for the PXF group and 0.36D ± 0.48 for the control group (p < 0.05). The MAE for the Hill-RBF 3.0 formula was 0.61D ± 0.84 for the PXF group and 0.42D ± 0.55 for the control group (p = 0.05). There were significant differences in MAE and MedAE between PXF group and control group measures (p < 0.05). In the PXF group there were no significant differences between the different formulae.
Conclusions
There were significant differences in accuracy of IOL power calculations in all formulae between PXF group and control group measures. PXF patients show hyperopic shift from predicted refraction. Barret universal II formula had the highest proportion of eyes with absolute error in prediction below or equal to 0.50 D in both PXF and control groups.
Publisher
Springer Science and Business Media LLC