Abstract
Abstract
Purpose
To determine between-method differences in corneal endothelial cell parameters using center and automated methods of non-contact specular microscopy (CellCheck software of Konan, Inc.) in glaucomatous eyes.
Methods
We analyzed the central corneal endothelial cell density (ECD) of 245 glaucomatous eyes using center (ECD-Ce) and automated methods (ECD-Au). Based on the ECD-Ce, we allocated subjects to Groups 1 to 10 (at 250 cells/mm2 intervals) and evaluated the ECD, coefficient of variation in cell area (CV), and percentage of hexagonal cells (HEX).
Results
There was a close correlation (r = 0.91) between the ECD values measured using both methods. However, ECD-Au were significantly higher than those measured by the center method when ECD-Ce was less than 2500 (in Groups 1 to 8; P < 0.001 to P = 0.006). The regression equation of (ECD-Au—ECD-Ce) = 1028–0.397*ECD-Ce shows greater deviation in eyes with lower ECD, and this difference became 0 when ECD -Ce was 2593 cells/mm2. None of the 44 subjects with an ECD-Ce of < 1000 cells/mm2 recorded an ECD-Au < 1000 cells/mm2. Compared with the center method, the automated method had higher and lower median CV and HEX values, respectively (P < 0.001). The between-method differences in both CV and HEX were negatively correlated with ECD-Ce (r = −0.49, P < 0.001 and r = −0.25, P < 0.001, respectively).
Conclusion
The automated method of the CellCheck software overestimates ECD in eyes with lower ECD values and may overlook risk of corneal decompensation.
Publisher
Springer Science and Business Media LLC
Reference15 articles.
1. Janson BJ, Alward WL, Kwon YH et al (2018) Glaucoma-associated corneal endothelial cell damage: a review. Surv Ophthalmol 63:500–506
2. Mannis MJ, Holland EJ (eds) (2017) Cornea, 4th edn. Rotterdam, Elsevier
3. McCarey BE, Edelhauser HF, Lynn MJ (2008) Review of corneal endothelial specular microscopy for FDA clinical trials of refractive procedures, surgical devices, and new intraocular drugs and solutions. Cornea 27:1–16
4. Konan Inc. KONAN MEDICAL [CellCheck Overview Video]. Available at: https://konanmedical.com/cellchek. Accessed February 22, 2020.
5. Huang J, Maram J, Tepelus TC et al (2018) Comparison of manual & automated analysis methods for corneal endothelial cell density measurements by specular microscopy. J Optom 11:182–191
Cited by
5 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献