Effectiveness and safety of VISULAS® green selective laser trabeculoplasty: a prospective, interventional multicenter clinical investigation

Author:

Pillunat Karin R.,Kretz Florian T. A.,Koinzer Stefan,Ehlken Christoph,Pillunat Lutz E.,Klabe Karsten

Abstract

Abstract Purpose To evaluate the effectiveness and safety of Selective Laser Trabeculoplasty (SLT) with the SLT mode of the VISULAS® green laser in patients with primary open-angle glaucoma (POAG). Methods This prospective, interventional multicenter clinical investigation included patients with POAG who either needed a treatment escalation because the individual intraocular pressure (IOP) target was not met or treatment initiation and had an IOP ≥ 17 mmHg at baseline in the study eye. The study was conducted in five research centers across Germany. Approximately 100 laser applications were delivered to 360° of the trabecular meshwork. Glaucoma medications were not modified during the 3-month follow-up to allow evaluation of the sole effect of VISULAS® green with SLT. Efficacy outcomes were postoperatively absolute and relative IOP changes at 1 and 3 months. Safety outcomes analyzed the rate of intra- and postoperative adverse events. Results Thirty-four eyes of 34 POAG patients were included. The overall mean number of preoperative glaucoma medications was 2.2 ± 1.4 in 29 treated eyes, 5 eyes were treatment naïve. Mean baseline IOP (mmHg) was 21.0 ± 2.69 and was reduced by − 3.53 ± 3.34 [95% CI − 4.61; − 2.45] and − 3.59 ± 3.41 [95% CI − 4.64; − 2.53] at the 1- and 3-month follow-up, respectively (p < 0.0001), with 48.5% of cases achieving a ≥ 20% IOP reduction at 3 months [95% CI = 30.8%; 66.5%]. The mean relative IOP reduction was − 16.4% and − 16.3% at 1 and 3 months, respectively (p < 0.0001). Potentially device- or procedure-related adverse events were mild to moderate and included 3 postoperative IOP-spikes and 6 reports regarding eye pain and discomfort. All were resolved without sequelae. Conclusions SLT performed with the VISULAS® green laser achieved clinically significant additional IOP reductions in medically treated as well as in treatment naïve eyes with POAG and there were no relevant safety issues. The results are comparable to other reported SLT studies.

Funder

Universitätsklinikum Carl Gustav Carus Dresden an der Technischen Universität Dresden

Publisher

Springer Science and Business Media LLC

Subject

Ophthalmology

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3