Evaluation of cannulated compression headless screws as an alternative implant for superior pubic ramus fracture fixation: a biomechanical study

Author:

Berk TillORCID,Zderic Ivan,Schwarzenberg Peter,Pastor Tatjana,Lesche Felix,Halvachizadeh Sascha,Richards R. Geoff,Gueorguiev Boyko,Pape Hans-Christoph

Abstract

Abstract Background/purpose Pubic ramus fractures account for the most common types of pelvic fractures. The standard surgical approach for superior pubic ramus fractures (SPRF) is a minimally invasive percutaneous screw fixation. However, percutaneous closed reduction and internal fixation of anterior pelvic ring injuries have high failure rates of up to 15%. The aim of this biomechanical study was to evaluate the stability of SPRF following stabilization with retrograde placed cannulated compression headless screw (CCHS) versus conventional fully and partially threaded screws in an artificial pelvic bone model. Methods SPRF type II as described by Nakatani et al. was created by means of osteotomies in eighteen anatomical composite hemi-pelvises. Specimens were stratified into three groups of six specimens each (n = 6) for fixation with either a 7.3 mm partially threaded cannulated screw (group RST), a 7.3 mm fully threaded cannulated screw (group RSV), or a 7.5 mm partially threaded cannulated CCHS (group CCS). Each hemi-pelvic specimen was tested in an inverted upright standing position under progressively increasing cyclic axial loading. The peak load, starting at 200 N, was monotonically increased at a rate of 0.1 N/cycle until 10 mm actuator displacement. Results Total and torsional displacement were associated with higher values for RST versus CCS and RSV, with significant differences between RST and CCS for both these parameters (p ≤ 0.033). The differences between RST and RSV were significant for total displacement (p = 0.020), and a trend toward significance for torsional displacement (p = 0.061) was observed. For both failure criteria 2 mm total displacement and 5° torsional displacement, CCS was associated with significantly higher number of cycles compared to RST (p ≤ 0.040). Conclusion CCHS fixation presented predominantly superior stability to the standard surgical treatment and could therefore be a possible alternative implant for retrograde SPRF screw fixation, whereas partially threaded screws in group RST were associated with inferior biomechanical stability.

Funder

University of Zurich

Publisher

Springer Science and Business Media LLC

Subject

Orthopedics and Sports Medicine,Surgery

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3