Author:
Liu Yanshi,Cai Feiyu,Liu Kai,Zhang Xingpeng,Li Hong,Fu Xuefei,Zhang Tao,Yusufu Aihemaitijiang
Abstract
Abstract
Purpose
As the monolateral external fixator is increasingly used in trauma-control and definitive management for high-energy long bone fractures, timing the fixator removal remains a challenge for surgeons. The purpose of this study was to determine the feasibility and effectiveness of the bony callus stiffness indirectly evaluated by the axial load-share ratio in vivo as a guide to removing a monolateral external fixator safely.
Methods
A total of 131 patients with tibial shaft fractures treated by the monolateral external fixator in our institution were collected from January 2013 to July 2019. In group I, the fixators were removed based on the clinical and radiological assessment only by the treating surgeon. As for group II, the axial load-share (LS) ratio test was accomplished by another medical team without the knowledge of the clinical results. The external fixator was removed when the mechanical test outcome (LS ratio < 10%) was consistent with the conclusion drawn from the clinical and radiological assessment (bone union achieved) by the treating surgeon.
Results
There was no statistical significance in demographic data between the two groups (P > 0.05). In group I, four patients suffered refracture (the refracture rate was 7.7%) after fixator removal and were successfully treated by an intramedullary nail. In group II, 71 patients underwent fixator removal after the first mechanical test, and another eight patients terminated the external fixation after the second test. None of the 79 patients in group II suffered refracture (the refracture rate was 0%). There was statistical significance in the refracture rate between the two groups (P < 0.05).
Conclusion
The bony callus stiffness indirectly evaluated by the axial load-share ratio in vivo using the additional circular frame components is an effectively quantitative indicator to complement the clinical assessment of fracture healing in a monolateral external fixation treatment. Removal of the monolateral external fixator is safe when the axial load-share ratio dropped below 10%.
Publisher
Springer Science and Business Media LLC
Subject
Orthopedics and Sports Medicine,Surgery
Cited by
5 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献