Abstract
AbstractA green, simple, quick and economical method is implemented for the first time for the simultaneous estimation of cetirizine (CTZ) and azelastine (AZE) as co-administered eye drops. The method relies on synchronous spectrofluorimetry with ∆λ = 60 nm. Cetirizine can be estimated at 231 nm and AZE can be measured at 294 nm, each at the other’s zero crossing point. All factors affecting the method were studied and properly optimized. Good correlation was obtained in the range of 0.1–2 µg mL−1 for both drugs. The limits of detection were 0.014 and 0.010 µg mL−1 and limits of quantitation were 0.043 and 0.029 µg mL−1 for CTZ and AZE, respectively. Moreover, ICH guidelines were carried out to validate the adopted method. The method was suitable for the analysis of CTZ and AZE in synthetic mixtures, eye drops and aqueous humor. The mean percentage of recoveries of CTZ and AZE in spiked aqueous humor were 99.83 and 99.37, respectively. Furthermore, Green Analytical Procedure Index (GAPI) and analytical Eco-scale approaches were used to evaluate the greenness of the suggested method.
Publisher
Springer Science and Business Media LLC
Subject
Law,Clinical Biochemistry,Spectroscopy,Sociology and Political Science,Social Sciences (miscellaneous),Clinical Psychology,Biochemistry