Author:
Silvis M. J. M.,Dekker M.,Zivelonghi C.,Agostoni P.,Stella P. R.,Doevendans P. A.,de Kleijn D. P. V.,van Kuijk J. P.,Leenders G. E.,Timmers L.
Abstract
Abstract
Background
Refractory angina is a growing and major health-care problem affecting millions of patients with coronary artery disease worldwide. The Coronary Sinus Reducer (CSR) is a device that may be considered for the relief of symptoms of refractory angina. It causes increased venous pressure leading to a dilatation of arterioles and reduced arterial vascular resistance in the sub-endocardium. This study describes the 5‑year Dutch experience regarding safety and efficacy of the CSR.
Methods
One hundred and thirty-two patients with refractory angina were treated with the CSR. The primary efficacy endpoint of the study was Canadian Cardiovascular Society (CCS) class improvement between baseline and 6‑month follow-up. The primary safety endpoint was successful CSR implantation in the absence of any device-related events.
Results
Eighty-five patients (67%) showed improvement of at least 1 CCS class and 43 patients (34%) of at least 2 classes. Mean CCS class improved from 3.17 ± 0.61 to 2.12 ± 1.07 after implantation (P < 0.001). The CSR was successfully implanted in 99% of the patients and only minor complications during implantation were reported.
Conclusion
The CSR is a simple, safe, and effective option for most patients with refractory angina. However, approximately thirty percent of the patients showed no benefit after implantation. Future studies should focus on the exact underlying mechanisms of action and reasons for non-response to better identify patients that could benefit most from this therapy.
Publisher
Springer Science and Business Media LLC
Subject
Cardiology and Cardiovascular Medicine
Cited by
18 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献