Author:
Oertel Markus F.,Stieglitz Lennart H.,Bozinov Oliver
Abstract
Abstract
Background
Magnetic resonance imaging (MRI)-guided laser interstitial thermal therapy (MRIgLITT) was demonstrated to be a viable neurosurgical tool. Apart from its variety of indications, different operative and technical nuances exist. In the present report, for the first time, the use and ability of a traditional Riechert-Mundinger (RM) stereotactic system combined with a novel drill guide kit for MRIgLITT are described.
Methods
A stereotactic frame-based setting was developed by combining an RM system with a drill guide kit and centering bone anchor screwing aid for application together with an MRIgLITT neuro-accessory kit and cooled laser applicator system. The apparatus was used for stereotactic biopsy and consecutive MRIgLITT with an intraoperative high-field MRI scanner in a brain tumor case.
Results
The feasibility of an RM stereotactic apparatus and a drill guide kit for MRIgLITT was successfully assessed. Both stereotactic biopsy and subsequent MRIgLITT in a neurooncological patient could easily and safely be performed. No technical problems or complications were observed.
Conclusion
The combination of a traditional RM stereotactic system, a new drill guide tool, and intraoperative high-field MRI provides neurosurgeons with the opportunity to reliably confirm the diagnosis by frame-based biopsy and allows for stable and accurate real-time MRIgLITT.