Abstract
Abstract
Background
Deep brain stimulation (DBS) is an established treatment for patients with medical refractory movement disorders with continuously increasing use also in other neurological and psychiatric diseases. Early and late complications can lead to revision surgeries with partial or complete DBS-system removal. In this study, we aimed to report on our experience with a frameless x-ray-based lead re-implantation technique after partial hardware removal or dysfunction of DBS-system, allowing the preservation of intracerebral trajectories.
Methods
We describe a surgical procedure with complete implant removal due to infection except for the intracranial part of the electrode and with non-stereotactic electrode re-implantation. A retrospective analysis of a patient series treated using this technique was performed and the surgical outcome was evaluated including radiological and clinical parameters.
Results
A total of 8 DBS-patients with lead re-implantation using the frameless x-ray-based method were enrolled in the study. A revision of 14 leads was performed, whereof a successful lead re-implantation could be achieved without any problems in 10 leads (71%). In two patients (one patient with dystonia and one patient with tremor), the procedure was not successful, so we placed both leads frame-based stereotactically.
Conclusions
The described x-ray-based technique allows a reliable frameless electrode re-implantation after infection and electrode dysfunction and might represent an efficient alternative to frame-based procedures for lead revision making the preservation of intracerebral trajectories possible.
Funder
Georg-August-Universität Göttingen
Publisher
Springer Science and Business Media LLC
Subject
Clinical Neurology,Surgery
Reference25 articles.
1. Bjerknes S, Skogseid IM, Saehle T, Dietrichs E, Toft M (2014) Surgical site infections after deep brain stimulation surgery: frequency, characteristics and management in a 10-year period. PLoS One 9:e105288
2. Blomstedt P, Bjartmarz H (2012) Intracerebral infections as a complication of deep brain stimulation. Stereotact Funct Neurosurg 90:92–96
3. Boviatsis EJ, Stavrinou LC, Themistocleous M, Kouyialis AT, Sakas DE (2010) Surgical and hardware complications of deep brain stimulation. A seven-year experience and review of the literature. Acta Neurochir 152:2053–2062