Abstract
Abstract
Background
Recently, a novel cryoballoon ablation catheter has demonstrated acute safety and efficacy in de novo pulmonary vein isolation (PVI) procedures in patients with paroxysmal atrial fibrillation (PAF). However, there are limited studies demonstrating the long-term efficacy. The aim of this study was to evaluate the long-term safety and efficacy of this novel cryoballoon in treating PAF.
Methods
This was a non-randomized, prospective, multicentre study enrolling 58 consecutive patients. Cryoablation was delivered for 180 s if time to isolation was ≤ 60 s. Otherwise a 240-s cryoablation was performed. One centre performed pre- and post-ablation high-density mapping (n = 9) to characterize lesion formation. After a 3-month blanking period, recurrence was defined as having any documented, symptomatic episode(s) of AF or atrial tachycardia. All patients were followed for 1 year.
Results
Acute PVI was achieved in 230 of 231 pulmonary veins (99.6%) with 5.3 ± 1.6 cryoablations per patient (1.3 ± 0.7 cryoablations per vein). Forty-three (77%) patients remained arrhythmia-free at 1-year follow-up. Four patients (6.9%) experienced phrenic nerve injury (3 resolved during the index procedure; 1 resolved at 6 months). One serious adverse device event was reported: femoral arterial embolism event occurring 2 weeks post-index procedure. For patients who underwent high-density mapping, cryoablation was antral with 50% of the posterior wall ablated.
Conclusions
Initial multicentre clinical experience with a novel cryoballoon has demonstrated safety and efficacy of PVI in patients with PAF. Ablation with this cryoballoon provides a wide, antral lesion set with significant debulking of the posterior wall of the left atrium.
Funder
Boston Scientific Corporation
Publisher
Springer Science and Business Media LLC
Subject
Physiology (medical),Cardiology and Cardiovascular Medicine
Cited by
7 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献