Introduction
Headspace gas chromatography–mass spectrometry (HS-GC–MS) is widely considered the gold standard of quantitative fecal VOC analysis. However, guidelines providing general recommendations for bioanalytical method application in research and clinical setting are lacking.
Objectives
To propose an evidence-based research protocol for fecal VOC analysis by HS-GC–MS, based on extensive testing of instrumental and sampling conditions on detection and quantification limits, linearity, accuracy and repeatability of VOC outcome.
Methods
The influence of the following variables were assessed: addition of different salt solutions, injection temperature, injection speed, injection volume, septum use, use of calibration curves and fecal sample mass. Ultimately, the optimal sample preparation was assessed using fecal samples from healthy preterm infants. Fecal VOC analysis in this specific population has potential as diagnostic biomarkers, but available amount of feces is limited here, so optimization of VOC extraction is of importance.
Results
We demonstrated that addition of lithium chloride enhanced the release of polar compounds (e.g. small alcohols) into the headspace. Second, a linear relationship between injection volume, speed and temperature, and fecal sample mass on the abundance of VOC was demonstrated. Furthermore, the use of a septum preserved 90% of the non-polar compounds. By application of optimal instrumental and sampling conditions, a maximum of 320 unique compounds consisting of 14 different chemical classes could be detected.
Conclusions
These findings may contribute to standardized analysis of fecal VOC by HS-GC–MS, facilitating future application of fecal VOC in clinical practice.
Publisher
Springer Science and Business Media LLC
Subject
Clinical Biochemistry,Biochemistry,Endocrinology, Diabetes and Metabolism
Reference17 articles.
1. Amann, A., Costello Bde, L., Miekisch, W., Schubert, J., Buszewski, B., Pleil, J., et al. (2014). The human volatilome: Volatile organic compounds (VOCs) in exhaled breath, skin emanations, urine, feces and saliva. Journal of Breath Research, 8, 034001. 2. Arasaradnam, R. P., Covington, J. A., Harmston, C., & Nwokolo, C. U. (2014). Review article: Next generation diagnostic modalities in gastroenterology–gas phase volatile compound biomarker detection. Alimentary Pharmacology & Therapeutics, 39, 780–789. 3. Berkhout, D. J., Benninga, M. A., van Stein, R. M., Brinkman, P., Niemarkt, H. J., de Boer, N. K., et al. (2016). Effects of sampling conditions and environmental factors on fecal volatile organic compound analysis by an electronic nose device. Sensors (Basel), 16, 1967. 4. Berkhout, D. J. C., Niemarkt, H. J., Buijck, M., van Weissenbruch, M. M., Brinkman, P., Benninga, M. A., et al. (2017). Detection of sepsis in preterm infants by fecal volatile organic compounds analysis: A proof of principle study. Journal of Pediatric Gastroenterology and Nutrition, 65, e47–e52. 5. Berkhout, D. J. C., van Keulen, B. J., Niemarkt, H. J., Bessem, J. R., de Boode, W. P., Cossey, V., et al. (2019). Late-onset sepsis in preterm infants can be detected preclinically by fecal volatile organic compound analysis: A prospective, multicenter cohort study. Clinical Infectious Diseases, 68, 70–77.
Cited by
13 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献
|
|