The influence of prolonged instrument manipulation on gas leakage through trocars
-
Published:2023-07-13
Issue:9
Volume:37
Page:7325-7335
-
ISSN:0930-2794
-
Container-title:Surgical Endoscopy
-
language:en
-
Short-container-title:Surg Endosc
Author:
Robertson DanielORCID, van Duijn Matthijs, Arezzo Alberto, Mintz Yoav, Boni Luigi, Baldari Ludovica, Carus Thomas, Chand Manish, Fuchs Hans, Ficuciello Fanny, Marconi Stefania, Mylonas George, Kim Young Woo, Nakajima Kiyokazu, Schijven Marlies, Valdastri Pietro, Sagiv Chen, Mascagni Pietro, Myśliwiec Piotr, Petz Wanda, Sánchez-Margallo Francisco, Horeman-Franse Tim,
Abstract
Abstract
Background
During laparoscopic surgery, CO2 insufflation gas could leak from the intra-abdominal cavity into the operating theater. Medical staff could therefore be exposed to hazardous substances present in leaked gas. Although previous studies have shown that leakage through trocars is a contributing factor, trocar performance over longer periods remains unclear. This study investigates the influence of prolonged instrument manipulation on gas leakage through trocars.
Methods
Twenty-five trocars with diameters ranging from 10 to 15 mm were included in the study. An experimental model was developed to facilitate instrument manipulation in a trocar under loading. The trocar was mounted to a custom airtight container insufflated with CO2 to a pressure of 15 mmHg, similar to clinical practice. A linear stage was used for prolonged instrument manipulation. At the same time, a fixed load was applied radially to the trocar cannula to mimic the reaction force of the abdominal wall. Gas leakage was measured before, after, and during instrument manipulation.
Results
After instrument manipulation, leakage rates per trocar varied between 0.0 and 5.58 L/min. No large differences were found between leakage rates before and after prolonged manipulation in static and dynamic measurements. However, the prolonged instrument manipulation did cause visible damage to two trocars and revealed unintended leakage pathways in others that can be related to production flaws.
Conclusion
Prolonged instrument manipulation did not increase gas leakage rates through trocars, despite damage to some individual trocars. Nevertheless, gas leakage through trocars occurs and is caused by different trocar-specific mechanisms and design issues.
Funder
Technische Universiteit Delft European Association for Endoscopic Surgery and other Interventional Techniques
Publisher
Springer Science and Business Media LLC
Reference38 articles.
1. Gurusamy KS, Vaughan J, Davidson BR (2014) Low pressure versus standard pressure pneumoperitoneum in laparoscopic cholecystectomy. Cochrane Database Syst Rev. https://doi.org/10.1002/14651858.CD006930.pub3 2. la Chapelle CF, Bemelman WA, Rademaker BMP, van Barneveld TA, Jansen FW (2012) A multidisciplinary evidence-based guideline for minimally invasive surgery Part 1: entry techniques and the pneumoperitoneum. Gynecol Surg 9:271–282. https://doi.org/10.1007/s10397-012-0731-y 3. Cahill RA, Dalli J, Khan M, Flood M, Nolan K (2020) Solving the problems of gas leakage at laparoscopy. Br J Surg 107:1401–1405. https://doi.org/10.1002/bjs.11977 4. Robertson D, Sterke F, van Weteringen W, Arezzo A, Mintz Y, Nickel F, Boni L, Baldari L, Carus T, Chand M, Fuchs H, Ficuciello F, Marconi S, Mylonas G, Kim YW, Nakajima K, Schijven M, Valdastri P, Sagiv C, Mascagni P, Myśliwiec P, Petz W, Sánchez-Margallo F, Horeman T (2022) Characterisation of trocar associated gas leaks during laparoscopic surgery. Surg Endosc 36:4542–4551. https://doi.org/10.1007/s00464-021-08807-1 5. Choi SH, Kwon TG, Chung SK, Kim TH (2014) Surgical smoke may be a biohazard to surgeons performing laparoscopic surgery. Surg Endosc 28:2374–2380. https://doi.org/10.1007/s00464-014-3472-3
|
|