Abstract
Abstract
Background
Improvements in laparoscopic partial nephrectomy (LPN) in order to minimize perioperative warm ischemia time (WIT), complications, and consequently patient outcome are desirable. Veriset™ is a ready-to-use hemostatic patch of absorbable oxidized cellulose and hydrogel components that has earlier been implemented in vascular and hepatic surgery. We report our experience using this device in LPN.
Methods
Patients with a solitary malignant renal mass suspicious for renal cancer underwent LPN with either the use of Veriset™ hemostatic patch (n = 40) or conventional suture technique (n = 40). Patient characteristics, operation time and WIT, postoperative course and complications were recorded retrospectively. Tumor complexity was calculated according to the R.E.N.A.L. score. Outcome was determined according to the “trifecta” criteria (negative surgical margin, WIT < 25 min, no complications within 30 days).
Results
No significant differences with regard to clinical parameters and median R.E.N.A.L. score (6) were observed between both groups. Operation time (mean 127.1 min vs. 162. 8 min; p = 0.001) and WIT were both lower in the Veriset™ group (14.6 min vs. 20.6 min; p = 0.01). No differences in surgical margins (p = 0.602) and overall complication rates at 30 (p = 0.599) and 90 days (p = 0.611) postoperatively were noticed. The surgical outcome according to “trifecta” was achieved in 65% of patients using Veriset™ and in 57.5% of patients by suture closure, respectively.
Conclusion
The hemostatic Veriset™ patch can successfully be implemented in LPN. Handling and application appear favorable, thereby reducing operation time and WIT. The present results suggest that the device may represent an alternative to parenchyma suturing in LPN.
Publisher
Springer Science and Business Media LLC
Cited by
5 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献