Purpose
The study aimed to construct a new retinal tack design with high retention forces to prevent spontaneous disentanglement in cases of complicated retinal surgery.
Methods
Six new forms for the peak of a retinal tack were developed using computer-aided design (CAD); then a prototype was produced for each model. Finally, standardised design testing was conducted using human (ex vivo) sclera by logging 15 consecutive measurements for each model.
Results
Seven different models underwent pull-out testing (six new models and the original tack model), but two tack models (Model 4, Model 5) failed to penetrate the human tissue. The highest pull-out forces (median) were measured for Model 3, followed by Model 6, Model 2 and Model 1. The original Heimann tack (Model H) was found to have the lowest retention forces.
Conclusion
The different tack designs altered the penetration and holding forces. The retention forces of the proposed peak design led to a significant increase in the retention forces that were more than twice as high as those in the original Heimann Model.