Purpose
The success of XEN Gel Stent (XEN) and Preserflo MicroShunt (Preserflo) implantation depends mainly on the development of bleb fibrosis. This study aimed to describe the histological findings of bleb fibrosis after XEN and Preserflo surgery.
Methods
This retrospective study included patients with different types of glaucoma who underwent revision surgery after XEN or Preserflo implantation. The available clinical information and histological samples of removed fibrotic tissue were analyzed.
Results
Thirty-six patients were included. Revision surgery was performed at a median of 195 (range = 31–1264) days after primary surgery. The mean intraocular pressure changed from 29.1 (± 10.3) mmHg at baseline to 18.3 (± 8.7) mmHg (− 37%; p < 0.0001) and 16.2 (± 4.2) mmHg (− 45%; p < 0.0001) after 6 and 12 months, respectively. Histological analysis revealed an increase in activated fibroblasts and macrophages in all specimens and a parallel orientation of fibroblasts in a minor part of the probe in 60% of the specimens. No pronounced inflammatory reaction in the form of lymphocytic or granulocytic infiltration was observed. The comparison of specimens from uveitic glaucoma and primary open-angle glaucoma patients revealed no significant differences.
Conclusions
The histological analysis of fibrotic blebs from the XEN and Preserflo implants did not show any pronounced immune or foreign-body reaction and revealed a similar histological pattern of failed blebs after trabeculectomy.