Abstract
AbstractCobalt intoxications from fractured hip endoprostheses have been described since the early 2000s. Typical symptoms include cardiomyopathy, neurological symptoms with visual and hearing loss and hypothyroidism. Less is known about long-term progression of these pathologies. This case report shows the long-term course of complications caused by cobalt and chromium exposure after substantial elimination of the source of intoxication. We report here a 63-year-old male Caucasian with severe cobalt and chromium intoxication. He presented 1 month after 2nd revision surgery of a broken hip endoprosthesis in a reduced general condition with signs of heart failure, pale skin and diminished hearing and vision. Blood analyses showed a cobalt concentration of 600 μg/L (reference value < 0.45 μg/L). Because the blood cobalt concentration decreased rapidly after surgery and symptoms improved, chelation therapy was not applied. Close clinical and toxicological monitoring was performed. The intoxication was not diagnosed until 6 years after the faulty hip joint revision and 3 years of clear signs of intoxication during a 2nd revision of the prosthesis. The patient’s ordeal could have been much shorter if his cobalt intoxication with neurologic, cardiac and thyroid symptoms had been detected earlier by toxicological blood tests. After the elimination of the source of poisoning, the long-term course showed constant excretion of cobalt and chromium over several years without chelation. Specific symptoms such as cardiomyopathy and neurological symptoms were declining. However, due to the continuous release of metal ions from the tissue, complete recovery did not occur.
Funder
Universitätsklinikum Hamburg-Eppendorf (UKE)
Publisher
Springer Science and Business Media LLC
Reference20 articles.
1. Steens W, Loehr JF, von Foerster G, Katzer A. Chronic cobalt poisoning in endoprosthetic replacement. Orthopade. 2006;35(8):860–4.
2. Savarino L, Greco M, Cenni E, Cavasinni L, Rotini R, Baldini N, Giunti A. Differences in ion release after ceramic-on-ceramic and metal-on-metal total hip replacement. Medium-term follow-up. J Bone Joint Surg Br. 2006;88(4):472–6.
3. Korovessis P, Petsinis G, Repanti M, Repantis T. Metallosis after contemporary metal-on-metal total hip arthroplasty. Five to nine-year follow-up. J Bone Joint Surg Am. 2006;88:1183–91.
4. Citak M, Gehrke T, Thieme O. An extreme case of systemic metallosis after implantation of a hip prosthesis. Dtsch Arztebl Int. 2018;115:862.
5. Aronson JK. Cobalt. In: Aronson JK, editor. Meyler’s Side Effects of Drugs. 16th ed. Waltham, MA: Elsevier; 2016. p. 490–1.
Cited by
3 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献