Abstract
Abstract
Objective
A volume coil with squared slots-end ring was developed to attain improved sensitivity for imaging of rat’s brain at 7 T.
Material and Methods
The principles of the high cavity resonator for the low-pass case and the law of Biot-Savart were used to derive a theoretical expression of $$B_1/i$$
B
1
/
i
. The slotted-end ring resonator showed a theoretical 2.22-fold sensitivity improvement over the standard birdcage coil with similar dimensions. Numerical studies were carried out for the electromagnetic fields and specific absorption rates for our coil and a birdcage coil loaded with a saline-filled spherical phantom and a digital brain of a rat.
Results
An improvement of the signal-to-noise ratio (SNR) can be observed for the slotted volume coil over the birdcage regardless of the load used in the electromagnetic simulations. The specific absorption rate simulations show a decrement for the digital brain and quite similar values with the saline solution phantom. Phantom and rat’s brain images were acquired at 7 T to prove the viability of the coil design. The experimental noise figure of our coil design was four times less than the standard birdcage with similar dimensions, which showed a 44.5% increase in experimental SNR.
Discussion
There is remarkable agreement among the theoretical, numerical and experimental sensitivity values, which all demonstrate that the coil performance for MR imaging of small rodents can be improved using slotted end-rings.
Funder
Consejo Nacional de Ciencia y Tecnología
Publisher
Springer Science and Business Media LLC
Subject
Radiology, Nuclear Medicine and imaging,Radiological and Ultrasound Technology,Biophysics