Abstract
Abstract
Objective
The excellent blood and fat suppression of stimulated echo acquisition mode (STEAM) can be combined with saturation recovery single-shot acquisition (SASHA) in a novel STEAM-SASHA sequence for right ventricular (RV) native T1 mapping.
Materials and methods
STEAM-SASHA splits magnetization preparation over two cardiac cycles, nulling blood signal and allowing fat signal to decay. Breath-hold T1 mapping was performed in a T1 phantom and twice in 10 volunteers using STEAM-SASHA and a modified Look-Locker sequence at peak systole at 3T. T1 was measured in 3 RV regions, the septum and left ventricle (LV).
Results
In phantoms, MOLLI under-estimated while STEAM-SASHA over-estimated T1, on average by 3.0% and 7.0% respectively, although at typical 3T myocardial T1 (T1 > 1200 ms) STEAM-SASHA was more accurate. In volunteers, T1 was higher using STEAM-SASHA than MOLLI in the LV and septum (p = 0.03, p = 0.006, respectively), but lower in RV regions (p > 0.05). Inter-study, inter-observer and intra-observer coefficients of variation in all regions were < 15%. Blood suppression was excellent with STEAM-SASHA and noise floor effects were minimal.
Discussion
STEAM-SASHA provides accurate and reproducible T1 in the RV with excellent blood and fat suppression. STEAM-SASHA has potential to provide new insights into pathological changes in the RV in future studies.
Publisher
Springer Science and Business Media LLC
Subject
Radiology, Nuclear Medicine and imaging,Radiological and Ultrasound Technology,Biophysics