Abstract
AbstractObjectiveTo review the major hardware components of low-field point-of-care MRI systems which affect the overall sensitivity.MethodsDesigns for the following components are reviewed and analyzed: magnet, RF coils, transmit/receive switches, preamplifiers, data acquisition system, and methods for grounding and mitigating electromagnetic interference.ResultsHigh homogeneity magnets can be produced in a variety of different designs including C- and H-shaped as well as Halbach arrays. Using Litz wire for RF coil designs enables unloadedQvalues of ~ 400 to be reached, with body loss representing about 35% of the total system resistance. There are a number of different schemes to tackle issues arising from the low coil bandwidth with respect to the imaging bandwidth. Finally, the effects of good RF shielding, proper electrical grounding, and effective electromagnetic interference reduction can lead to substantial increases in image signal-to-noise ratio.DiscussionThere are many different magnet and RF coil designs in the literature, and to enable meaningful comparisons and optimizations to be performed it would be very helpful to determine a standardized set of sensitivity measures, irrespective of design.
Funder
Horizon 2020 Framework Programme
Publisher
Springer Science and Business Media LLC
Subject
Radiology, Nuclear Medicine and imaging,Radiological and Ultrasound Technology,Biophysics
Cited by
8 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献