Abstract
AbstractThe expediency of replacing the corrugated ring of a vascular stent with a constant corrugation pitch by a corrugated ring of variable pitch is being investigated. To describe the change in pitch of corrugation, a special function is introduced. The equilibrium and deformation equations were analyzed using the asymptotic homogenization method. The dependence of the radial stiffness of a corrugated ring with a variable pitch of corrugation on the constructive parameters is obtained. The effectiveness of the design under consideration is evaluated by the area of the lumen provided by the vascular stent. Compared to stents with regularly corrugated rings, this area is larger which improves the ring stent efficiency.
Publisher
Springer Science and Business Media LLC
Cited by
2 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献