Author:
Wendling-Keim Danielle S.,Kren Elena,Muensterer Oliver,Lehner Markus
Abstract
Abstract
Purpose
Despite constantly improving developments in ventriculo-peritoneal shunt systems, most patients with hydrocephalus require revision or replacement at some point of time. Therefore, this study aimed to analyse parameters that are associated with shunt dysfunction.
Methods
In this retrospective study, we included 81 patients aged 0–17 who were treated at our institution. Demographic data, etiology of the hydrocephalus, type of valve implanted, reason for any revision procedures, any complications and survival time of the ventriculo-peritoneal shunts were detected. Statistical analysis was performed using SPSS. The significance level was set at p ≤ 0.05.
Results
Over a mean study period of 18 years, we analyzed 226 valves subjected to 146 revision operations in 81 patients. The etiology of the hydrocephalus (p = 0.874) and the age of the child at the time of VP shunt implantation (p = 0.308) did not have any impact on the shunt survival time. However, the type of the valve significantly changed the survival time of the shunt (p = 0.030). Pressure differential valves presented a longer survival time than gravitational valves.
Conclusion
The majority of patients in this study needed at least one replacement of the initial shunt system. Pressure differential valves may be beneficial for the survival time of the shunt system.
Funder
Universitätsklinik München
Publisher
Springer Science and Business Media LLC
Subject
General Medicine,Surgery,Pediatrics, Perinatology and Child Health
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献