Abstract
Abstract
Purpose
Roentgen stereophotogrammetric analysis is a technique to make accurate assessments of the relative position and orientation of bone structures and implants in vivo. While the precision and accuracy of stereophotogrammetry for hip and knee arthroplasty is well documented, there is insufficient knowledge of the technique’s precision and, especially accuracy when applied to rotational movements in the spinal region.
Methods
The motion of one cadaver lumbar spine segment (L3/L4) was analyzed in flexion–extension, lateral bending and internal rotation. The specific aim of this study was to examine the precision and accuracy of stereophotogrammetry in a controlled in vitro setting, taking the surrounding soft tissue into account. The second objective of this study was to investigate the effect of different focal spot values of X-ray tubes.
Results
Overall, the precision of flexion–extension measurements was found to be better when using a 0.6 mm focal spot value rather than 1.2 mm (± 0.056° and ± 0.153°; respectively), and accuracy was also slightly better for the 0.6 mm focal spot value compared to 1.2 mm (− 0.137° and − 0.170°; respectively). The best values for precision and accuracy were obtained in lateral bending for both 0.6 mm and 1.2 mm focal spot values (precision: ± 0.019° and ± 0.015°, respectively; accuracy: − 0.041° and − 0.035°).
Conclusion
In summary, the results suggest stereophotogrammetry to be a highly precise method to analyze motion of the lumbar spine. Since precision and accuracy are better than 0.2° for both focal spot values, the choice between these is of minor clinical relevance.
Funder
Medizinische Hochschule Hannover (MHH)
Publisher
Springer Science and Business Media LLC
Subject
Orthopedics and Sports Medicine,Surgery
Cited by
3 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献