Abstract
Abstract
Purpose
Treatment of pelvic organ prolapse (POP) often requires the use of synthetic mesh. In case of a novel and standardized bilateral apical fixation, both uterosacral ligaments are replaced by polyvinylidene-fluoride (PVDF) tapes. One of the main problems remains the fixation method, which should be stable, but also simple and quick to use. The current study evaluated biomechanical differences between the cervical tape fixation with sutures (group 1), non-absorbable tacks (group 2) and absorbable tacks (group 3) in an in vitro porcine model.
Methods
A total of 28 trials, conducted in three groups, were performed on porcine, fresh cadaver uteri. All trials were performed until mesh, tissue or fixation device failure occurred. Primary endpoints were the biomechanical properties maximum load (N), displacement at failure (mm) and stiffness (N/mm). The failure mode was a secondary endpoint.
Results
There was a significant difference between all three groups concerning the maximum load. Group 1 (sutures) supported a maximum load of 64 ± 15 N, group 2 (non-absorbable tacks) yielded 41 ± 10 N and group 3 (absorbable tacks) achieved 15 ± 8 N.
The most common failure mode was a mesh failure for group 1 and 2 and a fixation device failure for group 3.
Conclusion
The PVDF-tape fixation with sutures supports 1.5 times the load that is supported by non-absorbable tacks and 4.2 times the load that is supported by absorbable tacks. Nevertheless, there was also a stable fixation through tacks. Sutures are the significantly stronger and cheaper fixation device but may prolong the surgical time in contrast to the use of tacks.
Funder
Universitätsklinikum Köln
Publisher
Springer Science and Business Media LLC
Subject
Obstetrics and Gynecology,General Medicine
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献