Image quality evaluation of dual-layer spectral CT in comparison to single-layer CT in a reduced-dose setting

Author:

Do Thuy Duong,Rheinheimer Stephan,Kauczor Hans-Ulrich,Stiller Wolfram,Weber Tim,Skornitzke Stephan

Abstract

Abstract Objectives To quantitatively and qualitatively evaluate image quality in dual-layer CT (DLCT) compared to single-layer CT (SLCT) in the thorax, abdomen, and pelvis in a reduced-dose setting. Methods Intraindividual, retrospective comparisons were performed in 25 patients who received at least one acquisition of all three acquisition protocols SLCTlow (100 kVp), DLCThigh (120 kVp), and DLCTlow (120 kVp), all covering the venous-phase thorax, abdomen, and pelvis with matched CTDIvol between SLCTlow and DLCTlow. Reconstruction parameters were identical between all scans. Image quality was assessed quantitatively at 10 measurement locations in the thorax, abdomen, and pelvis by two independent observers, and subjectively with an intraindividual forced choice test between the three acquisitions. Dose-length product (DLP) and CTDIvol were extracted for dose comparison. Results Despite matched CTDIvol in acquisition protocols, CTDIvol and DLP were lower for SLCTlow compared to DLCTlow and DLCThigh (DLP 408.58, 444.68, 647.08 mGy·cm, respectively; p < 0.0004), as automated tube current modulation for DLCTlow reached the lower limit in the thorax (mean 66.1 mAs vs limit 65 mAs). Noise and CNR were comparable between SLCTlow and DLCTlow (p values, 0.29–0.51 and 0.05–0.20), but CT numbers were significantly higher for organs and vessels in the upper abdomen for SLCTlow compared to DLCTlow. DLCThigh had significantly better image quality (Noise and CNR). Subjective image quality was superior for DLCThigh, but no difference was found between SLCTlow and DLCTlow. Conclusions DLCTlow showed comparable image quality to SLCTlow, with the additional possibility of spectral post-processing. Further dose reduction seems possible by decreasing the lower limit of the tube current for the thorax. Key Points Clinical use of reduced-dose DLCT is feasible despite the required higher tube potential. DLCT with reduced dose shows comparable objective and subjective image quality to reduced-dose SLCT. Further dose reduction in the thorax might be possible by adjusting mAs thresholds.

Funder

Medizinischen Fakultät Heidelberg, Universität Heidelberg

Publisher

Springer Science and Business Media LLC

Subject

Radiology, Nuclear Medicine and imaging,General Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3