Evaluation of a single-breath-hold radial turbo-spin-echo sequence for T2 mapping of the liver at 3T

Author:

Bencikova Diana,Han Fei,Kannengieser Stephan,Raudner Marcus,Poetter-Lang Sarah,Bastati Nina,Reiter Gert,Ambros Raphael,Ba-Ssalamah Ahmed,Trattnig Siegfried,Krššák MartinORCID

Abstract

Abstract Objectives T2 mapping of the liver is a potential diagnostic tool, but conventional techniques are difficult to perform in clinical practice due to long scan time. We aimed to evaluate the accuracy of a prototype radial turbo-spin-echo (rTSE) sequence, optimized for multi-slice T2 mapping in the abdomen during one breath-hold at 3 T. Methods A multi-sample (fat: 0–35%) agarose phantom doped with MnCl2 and 80 subjects (73 patients undergoing abdomen MR examination and 7 healthy volunteers) were investigated. A radial turbo-spin-echo (rTSE) sequence with and without fat suppression, a Cartesian turbo-spin-echo (Cart-TSE) sequence, and a single-voxel multi-echo STEAM spectroscopy (HISTO) were performed in phantom, and fat-suppressed rTSE and HISTO sequences were performed in in vivo measurements. Two approaches were used to sample T2 values: manually selected circular ROIs and whole liver analysis with Gaussian mixture models (GMM). Results The rTSE-T2s values exhibited a strong correlation with Cart-TSE-T2s (R2 = 0.988) and with HISTO-T2s of water (R2 = 0.972) in phantom with an offset between rTSE and Cart-TSE maps (mean difference = 3.17 ± 1.18 ms). The application of fat suppression decreased T2 values, and the effect was directly proportional to the amount of fat. Measurements in patients yielded a linear relationship between rTSE- and HISTO-T2s (R2 = 0.546 and R2 = 0.580 for ROI and GMM, respectively). Conclusion The fat-suppressed rTSE sequence allows for fast and accurate determination of T2 values of the liver, and appears to be suitable for further large cohort studies. Key Points Radial turbo-spin-echo T2 mapping performs comparably to Cartesian TSE-T2 mapping, but an offset in values is observed in phantom measurements. Fat-suppressed radial turbo-spin-echo T2 mapping is consistent with T2 of water as assessed by MRS in phantom measurements. Fat-suppressed radial turbo-spin-echo sequence allows fast T2 mapping of the liver in a single breath-hold and is correlated with MRS-based T2 of water.

Funder

Medical University of Vienna

Publisher

Springer Science and Business Media LLC

Subject

Radiology, Nuclear Medicine and imaging,General Medicine

Cited by 6 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3