To shield or not to shield: shielding may have unintended effects on patient dose in CT

Author:

Larjava Heli Riitta SinikkaORCID,Eneh Chibuzor T. M.,Niiniviita Hannele M.

Abstract

Abstract Objectives The aim of the patient out-of-plane shield is to reduce the patient radiation dose. Its effect on tube current modulation was evaluated with the out-of-plane shield visible in the localizer but absent in the scan range in chest CT with different CT scanners. Methods An anthropomorphic phantom was scanned with six different CT scanners from three different vendors. The chest was first scanned without any shielding, and then with the out-of-plane shield within the localizer but outside the imaged volume. All pitch values of each scanner were used. The tube current values with and without the out-of-plane shield were collected and used to evaluate the effect of overscanning and tube current modulation (TCM) on patient radiation dose. Results The highest increase in cumulative mA was 217%, when the pitch was 1.531. The tube current value increased already 8.9 cm before the end of the scanned anatomy and the difference between the tube current of the last slices (with and without the out-of-plane shield in the localizer) was 976%. Conclusion Applying an out-of-plane shield outside the scanned volume but visible in the localizer images may increase the patient dose considerably if the scanner’s TCM function is based only on localizer images. Clinical relevance statement The use of an out-of-plane shield in CT may strongly increase the tube current modulation and thus provide the patient with a higher radiation dose. Key Points • Applying an out-of-plane shield outside the scanned volume but visible in the localizer images may increase patient radiation dose considerably. • The effect is visible with scanners that use solely localizer-based tube current modulation. • Features like overscanning may be difficult for the user to notice when planning the scanning, and yet they may affect tube current modulation and through it to patient dose.

Funder

Valtion erityisvastuualuerahoitus

University of Turku (UTU) including Turku University Central Hospital

Publisher

Springer Science and Business Media LLC

Subject

Radiology, Nuclear Medicine and imaging,General Medicine

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3