Abstract
Abstract
Background and Purpose
The use of flow diverters to treat aneurysms arising from small caliber parent vessels has been reported. This article reports the results of the first in experiences with the p48MW (p48 Movable Wire) in humans, a device specifically designed to target vessels 1.75–3 mm in diameter.
Methods
This monocentric study retrospectively reviewed the prospectively maintained database to identify all patients treated with the p48MW device between January 2017 and January 2019 at this institution. Patient demographics, aneurysm characteristics, angiographic and clinical follow-up were recorded as well as complications.
Results
A total of 25 patients (20 female) with an average age of 55 ± 12.9 years (range 34–84) with 25 aneurysms were identified. The majority of the aneurysms was located in the anterior circulation (19/25, 76%). The average aneurysm dome width was 3.98 ± 3.6 mm (range 1.2–13 mm). Complete occlusion was seen in 18/24 (75%) aneurysms with neck remnants in 1/24 (4.2%) and continued aneurysm filling seen in the remaining cases (5/24, 20.8%). Adequate occlusion was seen in 79.2% of aneurysms (Raymond Roy Classification [RRC] grade I or II) during the follow-up period. There was a single technical complication with inappropriate deployment of the first p48MW. There was a single clinical complication (4%); however, the patient made a complete recovery (modified Rankin Scale [mRS] 0) and one patient died secondary to uncontrollable status epilepticus following acute subarachnoid hemorrhage unrelated to the treatment.
Conclusion
The p48MW is safe and effective for the treatment of aneurysms including those arising from distal vessels.
Publisher
Springer Science and Business Media LLC
Subject
Neurology (clinical),Radiology, Nuclear Medicine and imaging
Cited by
26 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献