Comparison of a novel self-expanding transcatheter heart valve with two established devices for treatment of degenerated surgical aortic bioprostheses

Author:

Nikolayevska OlgaORCID,Conradi Lenard,Schirmer Johannes,Reichenspurner Hermann,Deuschl Florian,Blankenberg Stefan,Schäfer Ulrich

Abstract

Abstract Aims This study was performed to compare haemodynamic properties of a novel transcatheter heart valve (THV) with two established valve technologies for treatment of failing surgical aortic bioprosthetic valves (SAV). The ALLEGRA THV has been recently described with a proven safety and performance profile. Methods and results The study was designed as a retrospective, single-centre study investigating 112 patients (77.7 ± 7.1 years, 53.8% female, STS score 6.8 ± 5.8% and logEuroSCORE I 27.4 ± 16.1%) with failing SAV. Patients were treated with the ALLEGRA THV (NVT, n = 24), the CoreValve/EvolutR (MTD, n = 64) or the Edwards Sapien/Sapien XT/Sapien 3 (EDW, n = 24). Adverse events, haemodynamic outcomes and patient safety were analysed according to VARC-3 definitions. Overall procedural success was high (94.6%), even though 58.9% of the treated SAV were classified as small (true inner diameter < 21 mm). After treatment, the mean pressure gradient was significantly reduced (baseline: 33.7 ± 16.5 mmHg, discharge: 18.0 ± 7.1 mmHg), with a corresponding increase in effective orifice area (EOA). The complication rates did not differ in between groups. There was a trend to lower mean transvalvular gradients after implantation of self-expanding THV with supra-annular valve function, despite a higher frequency of smaller SAVs in the NVT and MTD group. Additionally, comparison between NVT and MTD revealed statistically lower transvalvular gradients (NVT 14.9 ± 5.0 mmHg, MTD 18.7 ± 7.5 mmHg, p = 0.0295) in a subgroup analysis. Conclusions Valve-in-valve (ViV) treatment of failing SAV with supra-annular design like the ALLEGRA THV resulted in favourable haemodynamic outcomes with similar low clinical event rates and may therefore be an interesting alternative for VIV TAVI. Graphical abstract

Funder

Universitätsklinikum Hamburg-Eppendorf (UKE)

Publisher

Springer Science and Business Media LLC

Subject

Cardiology and Cardiovascular Medicine,General Medicine

Reference36 articles.

1. Thourani VH, Suri RM, Gunter RL, Sheng S, O’Brien SM, Ailawadi G, Szeto WY, Dewey TM, Guyton RA, Bavaria JE, Babaliaros V, Gammie JS, Svensson L, Williams M, Badhwar V, Mack MJ (2015) Contemporary real-world outcomes of surgical aortic valve replacement in 141,905 low-risk, intermediate-risk, and high-risk patients. Ann Thorac Surg 99(1):55–61. https://doi.org/10.1016/j.athoracsur.2014.06.050

2. Beckmann A, Funkat AK, Lewandowski J, Frie M, Ernst M, Hekmat K, Schiller W, Gummert JF, Cremer JT (2015) Cardiac surgery in Germany during 2014: a report on Behalf of the German Society for Thoracic and Cardiovascular Surgery. Thorac Cardiovasc Surg 63(4):258–269. https://doi.org/10.1055/s-0035-1551676

3. Jones JM, O’Kane H, Gladstone DJ, Sarsam MA, Campalani G, MacGowan SW, Cleland J, Cran GW (2001) Repeat heart valve surgery: risk factors for operative mortality. J Thorac Cardiovasc Surg 122(5):913–918. https://doi.org/10.1067/mtc.2001.116470

4. David TE, Ivanov J, Armstrong S, Feindel CM, Cohen G (2001) Late results of heart valve replacement with the Hancock II bioprosthesis. J Thorac Cardiovasc Surg 121(2):268–277. https://doi.org/10.1067/mtc.2001.112208

5. Baumgartner H, Falk V, Bax JJ, De Bonis M, Hamm C, Holm PJ, Iung B, Lancellotti P, Lansac E, Rodriguez Munoz D, Rosenhek R, Sjogren J, Tornos Mas P, Vahanian A, Walther T, Wendler O, Windecker S, Zamorano JL, E. S. C. S. D. Group (2017) 2017 ESC/EACTS Guidelines for the management of valvular heart disease. Eur Heart J 38(36):2739–2791. https://doi.org/10.1093/eurheartj/ehx391