Abstract
Abstract
Background
The prospective WEARIT-II-EUROPE registry aimed to assess the value of the wearable cardioverter-defibrillator (WCD) prior to potential ICD implantation in patients with heart failure and reduced ejection fraction considered at risk of sudden arrhythmic death.
Methods and results
781 patients (77% men; mean age 59.3 ± 13.4 years) with heart failure and reduced left ventricular ejection fraction (LVEF) were consecutively enrolled. All patients received a WCD. Follow-up time for all patients was 12 months. Mean baseline LVEF was 26.9%. Mean WCD wearing time was 75 ± 47.7 days, mean daily WCD use 20.3 ± 4.6 h. WCD shocks terminated 13 VT/VF events in ten patients (1.3%). Two patients died during WCD prescription of non-arrhythmic cause. Mean LVEF increased from 26.9 to 36.3% at the end of WCD prescription (p < 0.01). After WCD use, ICDs were implanted in only 289 patients (37%). Forty patients (5.1%) died during follow-up. Five patients (1.7%) died with ICDs implanted, 33 patients (7%) had no ICD (no information on ICD in two patients). The majority of patients (75%) with the follow-up of 12 months after WCD prescription died from heart failure (15 patients) and non-cardiac death (15 patients). Only three patients (7%) died suddenly. In seven patients, the cause of death remained unknown.
Conclusions
Mortality after WCD prescription was mainly driven by heart failure and non-cardiovascular death. In patients with HFrEF and a potential risk of sudden arrhythmic death, WCD protected observation of LVEF progression and appraisal of competing risks of potential non-arrhythmic death may enable improved selection for beneficial ICD implantation.
Graphic abstract
Publisher
Springer Science and Business Media LLC
Subject
Cardiology and Cardiovascular Medicine,General Medicine
Cited by
18 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献