Abstract
Abstract
Background
The cryoballoon (CB) has become a standard tool for pulmonary vein isolation (PVI), but the technology is limited in certain ways. A novel RF-balloon (Heliostar™, Biosense Webster, CA, USA) promises the advantages of a balloon technology in combination with 3D mapping.
Methods
To assess procedural data and outcome, all patients undergoing RF-balloon PVI were included and compared with data from consecutive patients undergoing CB PVI for paroxysmal AF.
Results
A total of 254 patients (63 ± 13 years, 54% male) were included: 30 patients undergoing RF-balloon and 224 patients CB PVI. Baseline parameters were comparable. Procedure duration (104.3 ± 35.3 min vs. 69.9 ± 23.1 min; p ≤ 0.001) and fluoroscopy time (16.3 ± 7.1 min vs. 11.6 ± 4.9 min; p ≤ 0.001) were longer using the RF-balloon; ablation time (43.5 ± 17.9 vs. 36.4 ± 15.6; p = 0.08) did not differ, and time-to-isolation (TTI) was shorter (18.2 ± 7.0 s vs. 62.8 ± 35.1 s; p ≤ 0.001). Second-generation RF-balloon cases showed shorter ablation time and TTI at comparable procedure duration and fluoroscopy time. One pericardial effusion occurred with the RF-balloon due to complicated transseptal access. During CB PVI in 4/224 patients (1.8%), a phrenic nerve palsy was observed. After 12 months, 78% of patients after RF-balloon and 81% of patients after CB PVI (p = 0.5) were free from atrial arrhythmias.
Conclusion
The RF-balloon was safe and effective. Compared with the CB, TTI was shorter, but procedure durations and fluoroscopy times were longer. This can be attributed to a learning curve and the initial necessity for separate 3D map preparation. Considering the results with the second-generation RF-balloon, more experience is needed to determine the potential benefits.
Graphical Abstract
Publisher
Springer Science and Business Media LLC
Reference12 articles.
1. Hindricks G, Potpara T, Dagres N, Arbelo E, Bax JJ, Blomström-Lundqvist C et al (2020) ESC Guidelines for the diagnosis and management of atrial fibrillation developed in collaboration with the European Association for Cardio-Thoracic Surgery (EACTS). Eur Heart J 2020:1–126
2. Andrade JG, Wells GA, Deyell MW, Bennett M, Essebag V, Champagne J et al (2021) Cryoablation or drug therapy for initial treatment of atrial fibrillation. N Engl J Med 384:305–315
3. Wazni OM, Dandamudi G, Sood N, Hoyt R, Tyler J, Durrani S et al (2021) Cryoballoon ablation as initial therapy for atrial fibrillation. N Engl J Med 384:316–324
4. Kuck KH, Brugada J, Fürnkranz A, Metzner A, Ouyang F, Chun J et al (2016) Cryoballoon or radiofrequency ablation for paroxysmal atrial fibrillation. N Engl J Med 374:2235–2245
5. Fortuni F, Casula M, Sanzo A, Angelini F, Cornara S, Somaschini A et al (2020) Meta-analysis comparing cryoballoon versus radiofrequency as first ablation procedure for atrial fibrillation. Am J Cardiol 125:1170–1179