Abstract
AbstractIn vitro platforms such as bioreactors and microfluidic devices are commonly designed to engineer tissue models as well as to replicate the crosstalk between cells and microorganisms hosted in the human body. These systems promote nutrient supply and waste removal through culture medium recirculation; consequently, they intrinsically expose cellular structures to shear stress, be it a desired mechanical stimulus to drive the cell fate or a potential inhibitor for the model maturation. Assessing the impact of shear stress on cellular or microbial cultures thus represents a crucial step to define proper environmental conditions for in vitro models. In this light, the aim of this study was to develop a millifluidic device enabling to generate fully controlled shear stress profiles for quantitatively probing its influence on tissue or bacterial models, overcoming the limitations of previous reports proposing similar devices. Relying on this millifluidic tool, we present a systematic methodology to test how adherent cellular structures react to shear forces, which was applied to the case of microbial biofilms as a proof of concept. The results obtained suggest our approach as a suitable testbench to evaluate culture conditions in terms of shear stress faced by cells or microorganisms.Graphical Abstract