Abstract
Abstract
Better understanding of the lead curvature, movement and their spatial distribution may be beneficial in developing lead testing methods, guiding implantations and improving life expectancy of implanted leads.
Objective
The aim of this two-phase study was to develop and test a novel biplane cine-fluoroscopy-based method to evaluate input parameters for bending stress in leads based on their in vivo 3D motion using precisely determined spatial distributions of lead curvatures. Potential tensile, compressive or torque forces were not subjects of this study.
Methods
A method to measure lead curvature and curvature evolution was initially tested in a phantom study. In the second phase using this model 51 patients with implanted ICD leads were included. A biplane cine-fluoroscopy recording of the intracardiac region of the lead was performed. The lead centerline and its motion were reconstructed in 3D and used to define lead curvature and curvature changes. The maximum absolute curvature Cmax during a cardiac cycle, the maximum curvature amplitude Camp and the maximum curvature Cmax@amp at the location of Camp were calculated. These parameters can be used to characterize fatigue stress in a lead under cyclical bending.
Results
The medians of Camp and Cmax@amp were 0.18 cm−1 and 0.42 cm−1, respectively. The median location of Cmax was in the atrium whereas the median location of Camp occurred close to where the transit through the tricuspid valve can be assumed. Increased curvatures were found for higher slack grades.
Conclusion
Our results suggest that reconstruction of 3D ICD lead motion is feasible using biplane cine-fluoroscopy. Lead curvatures can be computed with high accuracy and the results can be implemented to improve lead design and testing.
Publisher
Springer Science and Business Media LLC
Subject
Cardiology and Cardiovascular Medicine,Biomedical Engineering
Reference18 articles.
1. Altman, P. A., J. M. Meagher, D. W. Walsh, and D. A. Hoffmann. Rotary bending fatigue of coils and wires used in cardiac lead design. J. Biomed. Mater. Res. 43:21–37, 1998.
2. Baxter, W. W., and A. D. McCulloch. In vivo finite element model-based image analysis of pacemaker lead mechanics. Med. Image. Anal. 5:255–270, 2001.
3. Baxter, W., J. Morissette, D. Roberts, and S. Schwartz. Acutely implanted cardiac lead shape measurement: physician prefrerence only? In: Summer Bioengineering Conference, 2003.
4. Baxter, W., N. Skadsberg, W. B. Johnson, G. Crossley, and B. Foreman. New unanticipatied insights on peak lead bending during pectoralis flexure. Heart Rhythm 7:309, 2010.
5. Cooke, D. J., A. Himes, and C. D. Swerdlow. Improved engineering standards for transvenous cardiac leads: a progress report from the Association for the Advancement of Medical Instrumentation Cardiac Rhythm Management Device Committee Leads Working Group. Heart Rhythm 16(6):958–959, 2019.