Abstract
AbstractThis work presents a comparative study for the analysis of carbohydrates for four common chromatographic methods, each coupled to mass spectrometry. Supercritical fluid chromatography (SFC), hydrophilic interaction liquid chromatography (HILIC), reversed-phase liquid chromatography (RP-LC) and gas chromatography (GC) with detection by triple quadrupole mass spectrometer (QqQ-MS) are compared. It is shown that gas chromatography and reversed-phase liquid chromatography, each after derivatisation, are superior to the other two methods in terms of separation performance. Furthermore, comparing the different working modes of the mass spectrometer, it can be determined that a targeted analysis, i.e. moving from full scan to single ion monitoring (SIM) and multiple reaction monitoring (MRM), results in an improvement in the sensitivity as well as the repeatability of the method, which has deficiencies especially in the analysis using HILIC. Overall, RP-LC–MS in MRM after derivatisation with 1-phenyl-3-methyl-5-pyrazolone (PMP) proved to be the most suitable method in terms of separation performance, sensitivity and repeatability for the analysis of monosaccharides. Detection limits in the nanomolar range were achieved, which corresponds to a mass concentration in the low µg/L range. The applicability of this method to different biological samples was investigated with various herbal liquors, pectins and a human glycoprotein.
Graphical abstract
Funder
Deutsche Forschungsgemeinschaft
Universität Duisburg-Essen
Publisher
Springer Science and Business Media LLC
Subject
Biochemistry,Analytical Chemistry
Cited by
11 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献