Abstract
Abstract
Purpose
We present a methodological overview of a respiration chamber at the Experimental and Clinical Research Center in Berlin, Germany. Since 2010, we investigated 750 healthy subjects and patients with various diseases. We routinely measure resting energy expenditure (REE), dietary-induced thermogenesis, and activity energy expenditure.
Methods
The chamber is a pull calorimeter with a total volume of 11,000 L. The majority of measurements is done with a flow rate of 120 L/min, yielding a favorable time constant of 1.53 h. The gas analysis system consists of two paramagnetic O2 sensors and two infrared CO2 sensors, one for incoming and one for outgoing air samples. O2 and CO2 sensors are calibrated simultaneously before each measurement with a 6 min calibration routine. To verify the accuracy of the whole the calorimetric system, it is validated every 2 weeks by 2 h acetone burning tests.
Results
Validation factors (calculated/measured) of 20 representative 2 h acetone burning tests were 1.03 ± 0.03 for $$\dot{V}{\text{O}}_{2}$$
V
˙
O
2
, 1.02 ± 0.02 for $$\dot{V}{\text{CO}}_{{2}}$$
V
˙
CO
2
, 0.99 ± 0.02 for RER, and 1.03 ± 0.03 for EE. Four repeated 60 min REE measurements of a healthy woman showed variabilities of 231.9 ± 4.8 ml/min for $$\dot{V}{\text{O}}_{2}$$
V
˙
O
2
(CV 2.1%), 166.0 ± 6.3 ml/min for $$\dot{V}{\text{CO}}_{{2}}$$
V
˙
CO
2
(CV 3.8%), 0.73 ± 0.03 for RER (CV 4.6%), and 4.55 ± 0.07 kJ/min for EE (CV 1.6%).
Conclusions
The data presented show that our respiration chamber produces precise and valid EE measurements with an exceptionally fast responsiveness.
Graphical Abstract
Funder
Charité - Universitätsmedizin Berlin
Publisher
Springer Science and Business Media LLC
Subject
Physiology (medical),Public Health, Environmental and Occupational Health,Orthopedics and Sports Medicine,General Medicine,Public Health, Environmental and Occupational Health,Physiology