Aseptic survival of the 1.5‐stage exchange arthroplasty for periprosthetic joint infection was acceptable when using an autoclaved femoral component and a new polyethylene insert

Author:

Song Sang Jun1,Hwang Sung Hyun1,Baek Hyun Jae1,Park Cheol Hee1ORCID

Affiliation:

1. Department of Orthopaedic Surgery Kyung Hee University College of Medicine Kyung Hee University Medical Center 26 Kyunghee‐daero, Dongdaemun‐gu 02447 Seoul Korea

Abstract

AbstractPurposeTo investigate the aseptic survival of 1.5‐stage exchange arthroplasty for periprosthetic joint infection (PJI) after total knee arthroplasty (TKA).MethodsEighty‐eight cases of 1.5‐stage exchange arthroplasty for PJI without reinfection were retrospectively analysed. The autoclaved femoral component and new polyethylene insert (PE) were implanted using antibiotic mixed cement. The explanted tibial component was not reinserted. The Western Ontario and McMaster Universities Osteoarthritis Index and range of motion were clinically evaluated preoperatively and at the last follow‐up (the last time for the implant in situ). Radiographically, hip–knee–ankle angle (HKA) and component positions were measured preoperatively, postoperatively (1 month after the 1.5‐stage exchange arthroplasty), and at the last follow‐up. The survival rate was analysed using the Kaplan–Meier method, in which failure was defined as reoperation due to aseptic failure. Mean period to failure and failure site were analysed. Factors affecting survival were investigated in terms of demographics and inappropriateness of the postoperative HKA (HKA > 0 ± 3°) and component positions (α angle > 95 ± 3°, β angle > 90 ± 3°, γ angle > 3 ± 3°, and δ angle > 87 ± 3°).ResultsThe spacer in‐situ time was 3.7 years (0.2–6.4). The clinical results improved hip–knee–ankle significantly at the last follow‐up. Radiographically, the average HKA was valgus 0.1° postoperatively. The average α, β, γ, and δ angles of the postoperative component positions were 95.9°, 90.4°, 3.8°, and 86.7°, respectively. The 1‐, 2‐, and 5‐year postoperative survival rates were 90.9%, 86.4%, and 80.6%, respectively. The mean period to failure was 2.0 years (0.2–5.3). There were 18 cases of aseptic loosening (20.8%), occurring on both the femur and tibial sides in 1 knee, and only on the tibial side in 17 knees. Inappropriate coronal position of the PE (β angle > 90 ± 3°) was a significant factor affecting survival (odds ratio = 5.491; p = 0.011).ConclusionThe aseptic survival of the 1.5‐stage exchange arthroplasty was acceptable when using an autoclaved femoral component and new PE. The appropriate coronal position of the PE helps ensure favourable survival of 1.5‐stage exchange arthroplasty.Level of evidenceIV

Publisher

Wiley

Subject

Orthopedics and Sports Medicine,Surgery

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3