Author:
Ohara Arisa,Machida Haruhiko,Shiga Hisae,Yamamura Wataru,Yokoyama Kenichi
Abstract
Abstract
Purpose
Ultrahigh-resolution CT (UHRCT) with slice collimation of 0.25 mm × 160 and matrix size of 1024 × 1024 has become clinically available. We compared the image quality of temporal bone CT (TBCT) between UHRCT and conventional multidetector CT (MDCT).
Materials and methods
We retrospectively enrolled 20 patients who underwent TBCT by MDCT (matrix size, 512 × 512) and subsequently by UHRCT (matrix size, 1024 × 1024). Two independent reviewers subjectively graded delineation of normal stapes, oval window, facial nerve canal, incudostapedial joint, and tympanic tegmen. We also quantified image noise in the cerebellar hemisphere. Between MDCT and UHRCT, we compared mean subjective grades using the Wilcoxon signed-rank test and the image noise using paired t test.
Results
Grades were significantly higher with UHRCT than with MDCT for all the anatomies (P < 0.001), whereas noise was significantly higher with UHRCT than with MDCT (P = 0.002).
Conclusion
For TBCT, UHRCT shows better delineation of the fine anatomical structures compared with MDCT.
Funder
Grant-in-Aid for Scientific Research (C) from the Japan Society for the Promotion of Science
Publisher
Springer Science and Business Media LLC
Subject
Radiology, Nuclear Medicine and imaging
Cited by
13 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献